Echo-doppler determination of the interatrial electromechanical conduction time. Comparison with the interatrial conduction time

Lanfranco Antonini, Salvatore Greco, Salvatore Toscano, Dennis Konopacki, Massimo Santini.
Department of Heart Diseases, San Filippo Neri Hospital, Rome, Italy

Abstract

The determination of the interatrial conduction time (IACT) is useful to optimize the AV interval of dual chambers pacemakers (DDD).
The IACT is determined using a bipolar transesophageal catheter. We propose the use of the interatrial electromechanical conduction time (IAEMCT) measured from the PM atrial spike to the beginning of the A wave of the transmitral flow by echo-doppler.
IACT and IAEMCT were compared in 36 pts with DDD.
There were no statistical significant differences between IACT and IAEMCT.
IAEMCT evaluates better the real AV synchronism since the electromechanical coupling is included.

Key Words

Pacing assessment – follow-up
DDD pacemakers, interatrial conduction time, interatrial electromechanical conduction time, Doppler echocardiography, Doppler mitral velocity wave, OA

 